TECH SOURCE
Guth Model 2100
Dom. Part No.: AS-2110-DR


The Guth Model 2100 was designed using a microcontroller, eliminating broad range mercury-in-glass thermostats. All components are chrome plated or stainless steel within the wet chamber area of the instrument. The all metal top housing is chrome plated brass, preventing rust and corrosion and provides for uniform headspace heating and mechanical security for all fasteners. The simulator utilizes a graduated laboratory glass container. The simulator has external fuse protection. The circuit boards have plug-in connectors. A calibration port is provided for insertion of a National Institute of Standards and Technology (NIST) reference mercury-in-glass thermometer to verify the proper operating temperature.

NHTSA APPROVED


PERFORMANCE
Temperature stability: ± .02°C
Accuracy of reading: ± .05°C
Warm-up time: <10 minutes*
Recovery time from test: <2 seconds
Liquid level 500 ml
*Ambient @20°C

SPECIFICATIONS
Construction: All metal, chrome plated
Weight: 5 lbs
Readout: Bright LED digital display
Sensor: Electronic Solid State
Power Input: 120 VAC, 60 cycle
220 - 240 VAC, 50 cycle
72 Watts










RS-232, 9600 Baud Digital Serial output of temperature & instrument status (polled)

IDLE MODE SHUTDOWN - Heater & motor turned off, idle mode displayed, controlled via RS-232 port

FEATURES
• Delivery tube heated for ethanol temperature stability
• Eight bit microprocessor control
• High stability internal voltage reference
• 12 bit digital accuracy from A/D converter
• Internal AC power transformer with AC connections isolated
• Software coded serial number
• Visual indication of system operation during warm-up (dot scan)
• Visual indication of heater activitiy via dot display
• Visual indication of system malfunction (Malfunction is displayed as an error number)
• Audible alarm indication of malfunction
• RFI protected


SYSTEM SHUTDOWN UNDER THE FOLLOWING CONDITIONS
• No solution in container
• Motor stops
• No output from shorted or open temperature sensor
• No change in solution temperature after initial 15 min. startup
• Temperature of solution <33.8°C / >34.2°C
• RFI present
• No reading from internal A/D converter

ERROR STATUS CODES
Error 1 = No solution or motor has stopped
Error 2 = Sensor shorted
Error 3 = Sensor open
Error 4 = Temperature >34.2°C
Error 5 = RFI present
Error 6 = Heater inoperative
Error 7 = Temperature <33.8°C
Error 8 = RS-232 error
Error 9 = Memory error

WARRANTY
One year, parts and labor
TECH SOURCE FORENSIC SERVICE AND SUPPLIES
1-800-329-0883